Table 2. Linear regression results (inflation number 2–50) for Otovent (A), sphere 1 (D), and sphere 2 (E)
Balloon n EquationCI interceptCI gradient P value of gradientPredicted number of inflationswhen maximum pressure is 40 mmHg
Otovent (A)8Y=–0.1775*X+86.6386.23 to 87.02–0.191 to –0.1641<0.0001262
Sphere 1 (D)10Y=–0.1108*X+77.1376.49 to 77.77–0.1323 to –0.0893<0.0001335
Sphere 2 (E)10Y=–0.2355*X+88.7088.05 to 89.34–0.2574 to –0.2136<0.0001206