Table 3. Correlation and agreement between measurement of PEF when standing and when sitting

Women

Men
All
Pearson’s correlation coefficient0.94 0.920.95
Bias between readings (slope, intercept)1.00 (1.00, 5.68)0.99 (1.04, −6.11)1.00 (1.03, −4.12)
Agreementa (95% CI)0.94 (0.912 to 0.959)0.92 (0.88 to 0.95)0.95 (0.938 to 0.964)
Limits of agreement (95% CI)5.81 (−53.60 to 65.22)12.39 (−71.51 to 96.28)8.72 (−62.66 to 80.10)
  • aLin’s concordance coefficient (Pearson’s correlation coefficient x bias between readings). PEF = peak expiratory flow.